4 diagnostic messages – Honeywell SMV 3000 User Manual

Page 143

Advertising
background image

1/99

SMV 3000 Transmitter User’s Manual

131

11.4

Diagnostic Messages,

Continued

Diagnostic Messages,
continued

Table 32

Non-Critical Status Diagnostic Message Table, Continued

SMV Status

SCT Status Message SFC Display Message TDC Status Message

Possible Cause

What to Do

9-7

Reynolds Number is Out

of Range

-

STATUS 9-7

The high or low Reynolds number
limit was exceeded.

• Verify high or low Reynolds

number limit.

• Calculate Reynolds number

for flow conditions causing the
message.

8-7

Sensor Mismatch PV3

SAVE/RESTORE

TYPE MISMATCH

SNSR MISMTCH PV3

Number of wires selected does not
match number of sensor wires
physically connected to the
transmitter.

Check sensor wiring and type.