4.3

Piping SMV 3000 Transmitter,

Continued

Installing flange
adapter,
continued

Table 6

Installing 1/2 inch NPT Flange Adapter

Step

Action

1

Insert filter screen (if supplied) into inlet cavity of process head.

2

Carefully seat Teflon (white) gasket into adapter groove.

3

Thread adapter onto 1/2-inch process pipe and align mounting holes
in adapter with holes in end of process head as required.

4

Secure adapter to process head by hand tightening 7/16-20 hex-head
bolts.

Example - Installing adapter on process head.

Process
Head

Filter Screen

Teflon Gasket

Flange Adapter

7/16 x 20 Bolts

21011

ATTENTION Apply an anti-seize compound on the stainless steel

bolts prior to threading them into the process head.

5

Evenly tighten adapter bolts to a torque of 47.5 to 54 N.m
(35 to 40 ft-lb).