34-ST-03-57

Page 5


Specifications

Operating Conditions – All Models

Parameter

Reference

Condition

(at zero

static)

Rated Condition

Operative Limits

Transportation

and Storage

°C °F

°C

°F

°C

°F

°C

°F

Ambient Temperature

25 ±1 77 ±2

-25 to 70

-13 to 158

-40 to 85

-40 to 185 -55 to 125 -67 to 257

Process Interface
Temperature

25 ±1 77 ±2

See Figure 13

-55 to 125 -67 to 257

Humidity %RH

10 to 55

0 to 100

0 to 100

0 to 100

Maximum Allowable
Working Pressure (MAWP)

MAWP is minimum of Body Rating or Seal Rating (See Model Selection Guide for Seal

MAWP)

Body MAWP

STR93D 750 psig (52 bar)
STR94G 500 psig (35 bar)

Vacuum Region, Minimum
Pressure - mmHg absolute

inH2O absolute


atmospheric
atmospheric

See Figure 13

Supply Voltage, Current,
and Load Resistance

Voltage Range: 10.8 to 42.4 Vdc at terminals
Current Range: 3.0 to 21.8 mA
Load Resistance: 0 to 1440 ohms (as shown in Figure 14)