3 specifications and environmental conditions 9, 1 specifications 9, 2 environmental conditions 9 – Fluke 9140 Dry-well Calibrator User Manual

Page 15: 3 specifications and environmental conditions, Specifications, Environmental conditions, 3specifications and environmental conditions, 1 specifications, 2 environmental conditions

Advertising
background image

3

Specifications and Environmental
Conditions

3.1

Specifications

The following table lists the specifications for this instrument. Accuracy speci-
fications are applicable for a one-year calibration interval. In line with normal
prudent metrology practices, Hart Scientific recommends a short-cycle interval
of six months for new units during the first year.

Power

115 VAC (±10%), 4.4 A or 230 VAC (±10%), 2.2 A, switchable 500 W

Ambient Temperature

5–50°C (41–122°F)

Operating Range

35–350°C (95–662°F)

Resolution

0.1°C or °F resolution

Readout

Switchable°C or °F

Accuracy

±0.5°C (±0.9°F), in holes greater than 1/4" (6.35 mm) accuracy is ±1°C
(±1.8°F)

Stability

±0.03°C at 50°C, ±0.05°C at 350°C

Uniformity

±0.4°C typical, ±0.1°C with similarly sized wells

Controller

Digital controller with data retention

Heater

500W

Heating Time

12 minutes from ambient to 350°C

Cooling Time

15 minutes from 350°C to 100°C

Cooling

2 speed internal fan

Stabilization Time

7 minutes

Immersion Depth

124 mm (4.88 inches)

Fault Protection

Sensor burnout and short protection, over temperature thermal cut-out, elec-
trical fuses

Test Wells

31.8 mm dia. x 123.7 mm deep (1.25" x 4.87"). Multi-hole inserts are
available.

Exterior Dimension

152.4 mm H x 85.7 mm W x 196.9 mm D (6" x 3.375" x 7.75")

Weight

2.7 kg (6 lbs.)

Fuse Rating

115 V: 6 A (fast acting), 250 V
230 V: 3.15 A (fast acting), 250 V

Safety

OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per
IEC1010-1

3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be

9

3 Specifications and Environmental Conditions

Specifications

Advertising