4 setting the temperature 12, Setting the temperature, 4 setting the temperature – Fluke 9140 Dry-well Calibrator User Manual

Page 18

Advertising
background image

dry-well will turn on and begin to heat to the previously programmed tempera-
ture set-point. The front panel LED display will indicate the actual dry-well
temperature.

4.4

Setting the Temperature

Section 7.2 explains in detail how to set the temperature set-point on the cali-
brator using the front panel keys. The procedure is summarized here.

(1) Press “SET” twice to access the set-point value.

(2) Press “UP” or “DOWN” to change the set-point value.

(3) Press “SET” to program in the new set-point.

(4) Press “EXIT” to return to the temperature display.

When the set-point temperature is changed the controller will switch the well
heater on or off to raise or lower the temperature. The displayed well tempera-
ture will gradually change until it reaches the set-point temperature. The well
may require 5 to 10 minutes to reach the set-point depending on the span. An-
other 5 to 10 minutes is required to stabilize within ±0.1°C of the set-point. Ul-
timate stability may take 15 to 20 minutes more of stabilization time.

9140 Dry-well Calibrator
User’s Guide

12