36

Command Description

Command
Format

Command
Example

Returned

Returned
Example

Acceptable
Values

Display Temperature

Read current set-point

s[etpoint]

s

set: 9999.99 {C or F}

set: 150.00 C

Set current set-point to n

s[etpoint]=n

s=350

Instrument
Range

Read temperature

t[emperature]

t

t: 9999.9 {C or F}

t: 55.6 C

Read temperature units

u[nits]

u

u: x

u: C

Set temperature units:

u[nits]=c/f

C or F

Set temperature units to Celsius

u[nits]=c

u=c

Set temperature units to
Fahrenheit

u[nits]=f

u=f

Read scan mode

sc[an]

sc

sc: {ON or OFF}

sc: ON

Set scan mode

sc[an]=on/off

sc=on

ON or OFF

Read scan rate

sr[ate]

sr

srat: 99.9 C/min

srat:12.4 C/min

Set scan rate

sr[ate]=n

sr=1.1

.1 to 99.9

Read hold

ho[ld]

ho

ho: open/closed, 99.9 {C
or F}

ho: open,
30.5 C

Secondary Menu

Read proportional band setting

pr[op-band]

pr

pb: 999.9

pb: 15.9

Set proportional band to

n

pr[op-band]=n

pr=8.83

Depends on
Configuration

Read heater power

(duty cycle)

po[wer]

po

po: 999.9

po: 6.5

Configuration Menu

Calibration Menu

Read R0 calibration parameter

r[0]

r

r0: 999.999

r0: 100.7

Set R0 calibration parameter to

n r[0]=n

r=100.7

97.0 to 105

Read ALPHA calibration
parameter

a[lpha]

al

al: 9.999999

al: 0.003865

Set ALPHA calibration parameter
to

n

a[lpha]=n

al=0.003865

.002 to .006

Read DELTA calibration
parameter

d[elta]

de

de: 1.50

Read DELTA calibration
parameter

d[elta]=n

de=1.37

de: 9.9999

0–3.0000

Operating Parameters Menu

Serial Interface Menu

Read serial sample setting

sa[mple]

sa

sa: 9

sa: 1

Set serial sampling setting to

n

seconds

sa[mple]=n

sa=0

0 to 999

Set serial duplex mode:

du[plex]=f[ull]/h[alf]

FULL or HALF

Set serial duplex mode to full

du[plex]=f[ull]

du=f

Set serial duplex mode to half

du[plex]=h[alf]

du=h

Table 2 9140 controller communications commands