Performance tests, Display test – Fluke 52 User Manual

Page 13

Advertising
background image

Thermometer

Performance Test and Calibration Required Test Equipment

5

Performance Test and Calibration Required Test Equipment

Refer to Table 2 for a list of test equipment used in the performance test and calibration
procedures.

Table 2. Recommended Test Equipment

Equipment

Required Characteristics

Recommended Model

DC Voltage Calibrator

Output Voltage: 0-1V
Accuracy: 0.0125 %
Resolution: 0.1

µ

V

Fluke 5700A or equivalent

Thermocouple connector

For K-type thermocouple

Fluke 80CK-M
Fluke P/N 779942

Thermocouple connector

Copper

Fluke P/N 601747

K-type Thermocouple wire

Standard

Omega

Dewar Flask/Cap

1- pint, or larger capacity

Thermos

Mercury Thermometer
(either

°

C or

°

F)

0.02

°

C Resolution

0.05

°

F Resolution

Princo Model ASTM56C
Princo Model ASTM56F

DB9 Adapter Connector

Fluke P/N 804549

IR Adapter

Fluke P/N 690518

PC with Microsoft Windows

®

installed

Performance Tests

Performance tests are recommended for incoming inspection, periodic maintenance, and
for checking the thermometer’s specifications. If the thermometer fails any part of the
performance tests, calibration and/or repair is indicated.

Before performing these tests:

1.

Allow the thermometer to stabilize to room temperature, +23

o

C

±

5

o

C (+73

o

F

±

9

o

F).

2.

Check the batteries and replace them if necessary.

Display Test

Turn the thermometer on while pressing the

h

button. All display segments are

displayed until the

h

button is released. Confirm that all the display segments are

showing as in Figure 1.