Fluke 52 User Manual

Page 16

Advertising
background image

51, 52, 53, 54 Series II

Service Manual

8

Table 3. Performance Test Values (K-Type Thermocouple)

Input

Voltage

Equivalent

Display Reading limits

Step

(mV DC)

Temperature

°

C

°

F

1

0.0000

0.0

°

C

32.0

°

F

-0.3 to 0.3

31.5 to 32.5

2

0.9193

23.0

°

C

73.4

°

F

22.7 to 23.3

72.9 to 73.9

3

-5.7297

-190.0

°

C

-310.0

°

F

-189.3 to -190.7

-308.9 to -311.1

4

40.8853

990.0

°

C

1814

°

F

989.2 to 990.8

1813 to 1815

5

48.8382

1200

°

C

2192

°

F

1199 to 1201

2190 to 2194

Verification of IR Communications Port (Fluke 53 and 54)

This section explains how to verify the operation of the IR Communications Port (IR
port). Perform the following procedure using a PC and HyperTerminal. The IR port can
also be verified using Flukeview

software.

To set up a HyperTerminal file, follow these steps:

1.

Connect the IR adapter cable to the DB9 Adapter connector. Connect the other end of
the DB9 Adapter to a COM port of the PC.

2.

Place the Fluke 53 or 54 in front of the IR adapter.

3.

Open up a HyperTerminal session. Select

Start button

,

Programs

,

Accessories

,

HyperTerminal

.

4.

Open up a

New Connection

with a name like “Fluke 54 IR Test”.

5.

When the

Connect To

box appears, select the COM1 or COM2 (the port that the IR

adapter cable is plugged into).

6.

Select

Port Settings

: 9600 baud rate, 8 data bits, no parity, 1 stop bit, and no flow

control.

7.

Under

File

, select

properties

. Verify that the correct port is selected. Click on

Configure

. Verify that port settings are correct (the baud rate may need to be

changed to 9600).

8.

Select the

Settings

tab then click on

ASCII Setup

. Check the following boxes:

Echo typed characters locally

Append line feeds to incoming line ends

Wrap lines that exceed terminal width

To open a HyperTerminal file to test the IR port, use the following procedure:

1.

Under

File

, select

Save As

. Use a file name such as “Fluke 54 IR Test” and then

click on

save

.

2.

Click on the

phone disconnect

icon.

3.

Click on

open folder

or under

File

select

Open

. Double click on Fluke 54 IR Test.

4.

Type in "ID" followed by the

Enter

key (notice the "ID" is capitalized).

5.

The display of the thermometer now shows the thermometer name (Fluke 53-II or
Fluke 54-II), and the units operating software version (e.g. Fluke 54-II, V1.01).

Advertising
This manual is related to the following products:

53, 51, 54