Calibration and offset adjustment, A/d calibration – Fluke 52 User Manual

Page 17

Advertising
background image

Thermometer

Calibration and Offset Adjustment

9

Calibration and Offset Adjustment

Calibration consists of two parts:

1.

Calibration of the A/D converter, using the set up shown in Figure 3.

2.

Reference junction calibration, using the ice bath test set up shown in Figure 2.

A/D Calibration

To calibrate the A/D portion of the thermometer, follow these steps:

1.

Connect the calibrator to the T1 input of the thermometer using copper (non-
thermocouple) wire and a copper thermocouple connector. Refer to Figure 3 for
proper equipment connections.

2.

Turn the thermometer on, with a small, blunt probe, press the calibration button
located on the back and in the center of the thermometer. Refer to Figure 4.

5700A
Calibrator

51, 52, 53, or 54
Series II
Thermometer

Copper
wire

Copper
Thermocouple
connector

+

-

_

- +

zb06f.eps

Figure 3. Calibration Connections