General information, Technical data, General data – YSI MultiLab IDS 4110 pH Electrode User Manual
Page 3: Measurement and application characteristics
 
4110
General information
3
ba76137e01
03/2013
General information
Automatic sensor
recognition
The sensor electronics with the stored sensor data is in the connecting head 
of the electrode. The data include, among other things, the sensor type and 
series number. With each calibration, the calibration data is written in the 
sensor and the calibration history is recorded. The data is recalled by the 
meter when the sensor is connected and is used for measurement and for 
measured value documentation. Storing the calibration data in the sensor 
ensures that the correct slope and asymmetry are automatically used if the 
sensor is operated with several meters.
The digital transmission technique guarantees the failure-free communi-
cation with the meter even with long connection cables. If the sensor firmware 
is enhanced by YSI, it can be updated via the meter.
Technical Data
General data
Measurement and application characteristics
Model
Reference 
elektrolyte
Junction
NTC
Special fatures
4110
Gel
Fiber
Yes
Plastic shaft
4110-3
Gel
Fiber
Yes
Plastic shaft
Model
pH measuring 
range
Allowed 
temperature 
range
Membrane 
resistance at 
25 °C
Typical 
application
4110
0.000 ... 14.000
0 ... 80 °C
< 1 GOhm
Field
4110-3
0.000 ... 14.000
0 ... 80 °C
< 1 GOhm
Field