Maintenance and cleaning

Cleaning

Remove water-soluble contamination by rinsing with deionized water. Other
types of contamination have to be removed as follows while the contact time
with the detergents should be kept as short as possible:

Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.

After cleaning

Rinse the electrode with deionized water and condition it in reference
electrolyte solution for at least 1 hour. Then recalibrate the electrode.

Wear parts and accessories

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household wash-
ing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Description

Model

Order no.

Reference electrolyte solution 250 ml
to fill the watering cap
(KCl 3 mol/l, Ag

+

free)

KCl-250

109 705Y