Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual
Page 41
Model 9060Z Manual
41
7.2.25 Communications Port Options
The 9060Z transmitter has a serial communications port available at terminals 18 to 22. The default
protocol is for RS‐232 running at 19,200 baud rate with 8 bits, Even parity and 1 stop bit.
Function
Options
60 Serial Interface
RS‐232 or RS‐485 (Use RS‐485 for MODBUS)
61 Serial Baud Rate
9600, 19200, 38400, 57600 (RS‐232 up to 19200 only)
62 Serial Parity
Even, Odd or No Parity
63
MODBUS
Address
Set the MODBUS address for this device between 1 and 246.
0 is to disable the MODBUS
7.2.26 Alarm Log Clearing
Every alarm that is instigated internally in the transmitter or as a process level is recorded in the alarm
log with the activation time, accepted time and the cleared time. The last 4000 events will be
recorded and then the oldest alarms will be dropped off as new ones occur. If you would like to delete
the log recording select CLEAR in SETUP step 64 by using the OPTION keys and press the ENTER key.
The message “Alarm Log Cleared” will be displayed.