3.3.11 Signal Characterizer

This function is used to compensate the output

for non-linear applications. The characterized

values are applied to the 4-20 mA output. For

the measured pressure, a maximum of nine

coordinates can be specified between 0-100%.

Perform the coordinate settings while the

S.C. at

S.C. menu parameter is “Disabled”.
To apply the settings to the output, set the

S.C.

parameter to “Enabled”.
Note that the transmitter rejects the activation of the

function by AL. 60 with the following transmitter’s

status:

• When the specified coordinates of x and y are

not incremental as the input increases.

• When the output mode of the output signal is

set as “Sq root”; at the same time, the low cut

mode is set to “Linear”.

F0308.ai

Y

X

100%

0%

100%

INPUT

OUTPUT

Input pressure in %

Characterized value

Follow the steps below to perform the signal

characterizer.
<1> Set the desired number of coordinates on the

line graph

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Detailed setup →

Signal condition → S.C. menu →

EJX_HART 5[1.2]

DTM

Configuration → Signal

Characterizer Menu →

→ Num of points

Set the number between 0 and 9

<2> Set the coordinates

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Detailed setup →

Signal condition → S.C. menu →

EJX_HART 5[1.2]

DTM

Configuration → Signal

Characterizer Menu →

→ Point setting

Set the coordinates (X-axis,

Y-axis)

<3> Apply the settings

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Detailed setup →

Signal condition → S.C. menu →

EJX_HART 5[1.2]

DTM

Configuration → Signal

Characterizer Menu →

→ S.C.

Select “Enabled” or “Disabled”

3.3.12 Alarm

The function is used to display the alarm codes

when the input pressure exceeds the specified

value within the calibration range. The same is

available for the input static pressure and the

capsule temperature on the pressure sensor.

Refer to table 4.5 Alarm Message Summary for the

specific alarm code to be generated.

(1) Alarm Setting
Select the process variable at

Process Alert which

the alarm is set, then set the alert mode for that

value.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Detailed setup

→ Output condition → Process

Alerts →

EJX_HART 5[1.2]

DTM

Configuration → Process Alerts →

Selection of the

process variable

for alarm

→ Pres Alert (DTM only) →

Pres Alert mode: Pressure
→ SP Alert (DTM only) →

SP Alert mode: Static pressure
→ Temp Alert (DTM only) → Temp

Alert mode: Capsule temperature

Selection of alert

mode

Off: Disable the alert function
Hi. Al Detect: High side alert

detection
Lo. Al Detect: Low side alert

detection
Hi/Lo. Al Detect: High and Low

side alert detection