4.2.2.1 Blockage Detection

Limit parameter

When the parameter based on pressure fluctuation

exceeds the preset value, EJX diagnoses an

impulse line as blockage and gives an alarm. The

threshold values are set to Limit parameter shown

in below table.

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Diag/Service

→ Diag Parameters → ILBD

Parameters → Configuration →

Diag Lim →

EJX_HART 5[1.2]

DTM

Diag and Service → Advanced Diag

Configurations → Impulse Line

Blockage Detection → Threshold

→ Sensitivity →

Table 4.2

Limit Parameter

#

Parameter

Threshold value

[1] Lim fDPmax

Threshold to detect “A

Blocking” by using

Ratio fDP

[2] Lim fDPmin

Threshold to detect “B

Blocking” by using

Ratio fDP

[3] Lim fSPlmax

Threshold to detect “Large

Fluct L” by using

Ratio fSPl

[4] Lim fSPlmin

Threshold to detect “L Side

Blocking” by using

Ratio fSPl

[5] Lim fSPhmax

Threshold to detect “Large

Fluct H” by using

Ratio fSPh

[6] Lim fSPhmin

Threshold to detect “H Side

Blocking” by using

Ratio fSPh

[7] Lim BlkFmax

Threshold to detect “H Side

Blocking” by using

BlkF

[8] Lim BlkFmin

Threshold to detect “L Side

Blocking” by using

BlkF

[9] Lim DPAvgmax Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

[10] Lim DPAvgmin

Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

Table 4.3 shows the default values at the factory

setting.

NOTE

• When ILBD is performed for the first time,

use the default value. If the pressure

fluctuation amplitude is low or a false alarm

is often generated after ILBD is performed,

change the values of Limit parameters

according to the procedure described in

subsection 4.2.2.10. Tuning