<4. Diagnostics>

4-18

IM 01C25T01-06EN

Combination of threshold for sensitivity parameter

High

Medium

Low

Lim fDPmax

1.50

3.00

3.00

Lim fDPmin

0.40

0.30

0.20

Lim fSPlmax

5.00

5.00

5.00

Lim fSPlmin

0.50

0.50

0.30

Lim fSPhmax

5.00

5.00

5.00

Lim fSPhmin

0.50

0.50

0.30

Lim BlkFmax

0.60

0.60

0.80

Lim BlkFmin

-0.60

-0.60

-0.80

Sampling Period

If fluctuating around the threshold value, an

alarm maybe often generated. When the above

phenomenon happens, the sampling time (

Diag

Period) can be changed so as to enhance the

accuracy of the blockage detection.
The longer the sampling time, better the expected

accuracy.

(1) Set “Stop” to

Diag Mode.

(2) Enter the value to

Diag Period within the

range of 20 to 65535 (seconds).

Note: Set to “Calculation” after setting the parameter.

Also, the accuracy can be improved by increasing

the number of

Diag Supp Count.

ILBD Range Setting

If flow/differential pressure is less than the default

threshold value of

Lim DPAvgmin, pressure

fluctuation is not large enough to detect the

blockage. To prevent the fault blockage detection,

the threshold should be changed to larger value.

(1) Set “Stop” to

Diag Mode.

(2) Enter the value to

Lim DPAvgmin or Lim

DPAvgmax.

Note Set to “Calculation” after setting the parameter.

• Procedure to call up the

threshold display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Diag/Service

→ Diag Parameters → ILBD

parameters → Configuration →

Diag Lim →

EJX_HART 5[1.2]

DTM

Diag and Service → Advanced

Diag Configuration → Impulse Line

Blockage Detection → Threshold →

Ratio fDP Compensation

When the flow change is too large or small, an

alarm maybe often generated. When the above

case happens, the Ratio fDP can be compensated

so as to enhance the accuracy of the blockage

detection.

Diag DPComp

When “Compensation” is selected in

Diag

DPComp, Ratio fDP is compensated by following

formula and used as treatable monitoring value,

CRatio fDP.

=

CRatio fDP

CRatio fDP

Ref fDP

fDP

DPAvg

Ref DPAvg

On the other hand, if the compensation is not

necessary, “Non-compensation” is selected in

Diag

DPComp and Ratio fDP is used as NRatio fDP.

=

NRatio fDP

Ref fDP

fDP

• Procedure to call up the display

DD and DTM

(excluding EJX_

HART 5[1.2])

[Root Menu] → Diag/Service

→ Diag Parameters → ILBD

parameters → Status → Diag

Vriables → CRatio fDP/NRatio fDP

EJX_HART 5[1.2]

DTM

Diag and Service → Advanced Diag

Variables → CRatio fDP/NRatio fDP

4.2.2.11 Reset of Reference Value

When there are large flow change or the change of

fluid conditions, obtain the reference value again. If

flow change by a comparison of the reference value

is ±25% or more, obtain the reference value again.