1 operating instructions, 1 using this manual, 2 product description – Eppendorf Self Test Function for Mastercycler nexus User Manual

Page 4: 1 delivery package, 2 features, 3 operation, 1 activating the self-test function, Operating instructions 1.1 using this manual, Operation 3.1 activating the self-test function

Advertising
background image

Self Test Function for Mastercycler® pro and Mastercycler® nexus

4

1

Operating instructions

1

Operating instructions

1.1 Using this manual

These instructions for use for the self-test function of the Mastercycler pro and Mastercycler

nexus is a supplement to the Thermocycler operating instructions and does not replace the

Thermocycler instructions.
Therefore, the Thermocycler operating instructions must be read before using the self-test

function for the first time. The current version of the instructions is available on the internet at

www.eppendorf.com

2

Product description

2 Product description

2.1 Delivery package

2.2 Features

The self-test function can be used with the following devices:

•

Mastercycler pro family of devices and the 6320 control panel (with USB ports)

•

Mastercycler nexus family of devices

Eliminating the need for an external measuring system, the self-test function is used to quickly

check of the following thermoblock characteristics:

•

Heating and cooling rates

•

Correct function of the temperature control circuit

•

Temperature homogeneity within each of the three block zones (left, center and right zone)

•

Temperature homogeneity across the entire block

A certificate can be output as a PDF after the self-test has been completed successfully.
The self-test USB key also activates the self-test function for all connected Mastercyclers.

3

Operation

3 Operation

3.1 Activating the self-test function

Requirement

•

A user is logged on (if the

PIN

function is activated).

•

The navigation tree is displayed.

Connect the self-test USB key to an open USB port on the Thermocycler.
The message

Please wait

appears. The self-test USB key is detected after a few seconds.

The self-test function is now unlocked.

Quantity Order No.

(International)

Order No.

(North America)

Description

1

6320 071.001

950030040

Selftest USB Key

For Mastercycler pro and Mastercycler nexus

1

6320 071.168

-

Mastercycler pro Self Test Function -

Instructions for Use

Advertising