7 - 8

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

6

Highlight default.tun. Click OK. See

Figure 7-12

.

NOTE: This will overwrite the

default.tun

file. If an overwrite is not

desired, see

section 10.3 on page 10-11

for information on this subject.

Figure 7-12 Save "default.tun"

7

Replace the existing

default.tun

file when prompted.

8

Reboot HAPSITE from the front panel MAIN menu. Refer to

section 3.8, EXIT

Menu, on page 3-59

for more information on how to reboot the HAPSITE.

9

If the HAPSITE fails to tune after steps 1-9 have been performed. See

section

7.6, Performing Manual Tune, on page 7-25

for information on performing a