Sensor Tooling (%) pane

Compensates for differences due to sensor and substrate geometry of the
Thickness measured by the sensor and the actual Thickness of material deposited
on the substrate (see

section 4.3 on page 4-2

). A value of 0.0 makes a sensor

unavailable.

Sensor 1 to 6 (IQM-233) . . . 0.0 to 999.0%

Sensor 1 to 8 (SQM-242) . . 0.0 to 999.0%

2.3.3.5.11 Errors Tab

Error detection and the action to occur upon the error are configured in the Errors
tab. See

Figure 2-32

.

Figure 2-32 Film Edit window - Errors tab

HINT: Until the stability and repeatability of the Process has been established, the

Enabled check boxes for Control Error, Crystal Quality, and Crystal
Stability
should be cleared, and the check box for Crystal Fail be selected.

On Error pane

The selected action is initiated if an error condition occurs as determined by the
Control Error pane selections. The possible actions are:

Ignore . . . . . . . . . . . . . . . . . All error conditions are ignored. The PID loop

attempts to control the rate until the deposition
phase is completed.