IP

N 07

4-

58

5-

P1

A

IQS-233 Operating Manual

Physical Sensor # . . . . . . . None, 1 to 6 (IQM-233 cards)

None, 1 to 8 (SQM-242 cards)

The Physical Sensor numbers 1 to 3 or 1 to 4
correspond to the same numbered sensor
connection on the IQM-233 or SQM-242 card,
respectively. The Physical Sensor numbers 4 to 6 or
5 to 8 correspond to the sensor connections 1 to 3 or
1 to 4 on the second IQM-233 or SQM-242 card,
respectively. Refer to the IQM-233 or SQM-242
operating manual for information about the sensor
connections.

NOTE: When two cards are installed, computers

may vary in which PCI express slot
corresponds to card 1 (Physical Sensor
numbers 1 to 3 or 1 to 4).

Dual . . . . . . . . . . . . . . . . . . Select the Dual check box to select the

corresponding sensor pair as the primary and
secondary channels of a dual sensor. If a crystal fail
occurs for the primary channel, the dual sensor
shutter exposes the secondary crystal and the
secondary channel is used for measurement. A PLC
is required to provide the relay for the dual sensor
shutter (see

Chapter 3, Digital I/O

).

Monitor Output . . . . . . . . . . None, Output 1 to 6 (IQM-233 cards)

None, Output 1 to 4 (SQM-242 cards)

Select the name of an Output to map to the
corresponding sensor. Multiple sensors may be
mapped to the same Output name for sensor rate
averaging.