Appearance of hydraulic fluid sample – MTS Series 505G2 SilentFlo Hydraulic Power Unit Model 505G2-180 User Manual

Page 89

Advertising
background image

7. If there still is any uncertainty regarding fluid quality, obtain another sample of the fluid and have it analyzed.

The fluid tests should include chemical analysis, particle count, and viscosity checks. Most oil companies
have facilities for performing these tests, or consult MTS Systems Corporation.

8. Remove the hose and valve assembly.

Appearance of Hydraulic Fluid Sample

What to Do

Problem

Fluid Properties

Analyze the fluid and replace as
required by the analysis.

Indicates chemical breakdown or
the hydraulic fluid has been

Dark colored fluid

allowed to rise above the

If the cause was overheating, flush
the entire system to remove varnish

maximum recommended
temperature (severe overheating)

or residue. Consult your MTS Field
Service Engineer to determine if
changes or adjustments to your
hydraulic system are required.

Analyze the fluid and replace as
required by the analysis.

Indicates chemical breakdown or
the hydraulic fluid has been

Burnt odor

allowed to rise above the

If the cause was overheating, flush
the entire system to remove varnish

maximum recommended
temperature (severe overheating)

or residue. Consult your MTS Field
Service Engineer to determine if
changes or adjustments to your
hydraulic system are required.

Replace the fluid

Indicates chemical breakdown

Opaque fluid

Check the heat exchanger for
damage.

Indicates water is present in the
fluid

Milky appearance

Look for other water sources if the
water does not appear to be coming
from the heat exchanger.

Identify and correct the source of the
water leakage and replace the fluid if
necessary.

Analyze the fluid and replace all filter
elements.

Indicates collapsed, ruptured, or
clogged filter(s)

Sediment at the bottom of
the sample container (after
sample has stood overnight)

Clean the fluid or replace it as
required by the analysis.

Series 505G2 SilentFlo™ Hydraulic Power Unit Product Information 89

Maintenance