Setting low pressure – MTS Hydraulic Service Manifold Model 293-12B User Manual

Page 44

Advertising
background image

Description

Item

Main Filter Indicator

1

6

A pop-up indicator is activated when the main pressure filter element becomes dirty.

Pilot Filter Indicator

2

7

A visual indicator is activated when the pilot pressure filter element becomes dirty.

Setting Low Pressure

The pressure delivered in high-pressure mode is determined by the HPU output pressure. Refer to the
operation section of the appropriate HPU manual for instructions on setting this level. In low-pressure
mode, the pressure delivered is determined by the setting of the low-pressure adjustment in the HSM.

Refer to the appropriate section for the location of the low-pressure adjustment. The setting can be changed
by using the following procedure.

1. Turn the HPU to low pressure.

2. Turn the HPU to high pressure.

3. Activate the HSM in low-pressure mode. Use the HSM pressure gage to check the present low-pressure

setting.

4. While holding the setting of the low-pressure adjustment with a hex key, turn the locking nut 1/2 turn

counterclockwise.

5. Set the pressure to the desired level, as registered on the gage. Clockwise rotation increases pressure.

6. While holding the setting of the low-pressure adjustment with the hex key, tighten the locking nut 1/2

turn clockwise.

7. Check the pressure at the gage. If necessary, repeat Steps 4 through 6 to improve the accuracy of the

setting.

8. Turn off the system.

1

If you start the HPU and hydraulic fluid is cold, the filter indicator can pop out. When this happens: allow
the fluid to warm up, press the filter indicator, and replace the filter element only if the indicator pops out
again.

2

If you start the HPU and hydraulic fluid is cold, the filter indicator can pop out. When this happens: allow
the fluid to warm up, press the filter indicator, and replace the filter element only if the indicator pops out
again.

44 | Series 293 Hydraulic Service Manifold Product Information