Tools and equipment, Calibration method, Tools and equipment -2 calibration method -2 – Rockwell Automation 1797-IRT8 FLEX Ex Thermocouple/RTD/mV Input Module User Manual User Manual

Page 54

Advertising
background image

Publication 1797-6.5.2 - February 2001

5-2 Calibrating Your Module

Tools and Equipment

To calibrate your analog I/O modules, you will need the following
tools and equipment:

Calibration Method

Calibration is done through the I/O Configuration Software which is
part of your programming software package, such as RSLogix 5.

Select the Calibration tab in the software and follow the instructions
for complete module calibration..

Tool or Equipment:

Description:

Precision Resistors

High precision resistors:
383W, 0.01%, 5ppm/

o

C

100W, 0.01%, 5ppm/

o

C

10kW, 0.5%, 5ppm/

o

C

Precision Voltage Source

+320mV, 1mV resolution

PC and Interconnect Cable

Programming terminal for A-B family processors

Advertising