Manual Supplement

QED-6 Users

2 11/10

On page 2-21, under step 4, replace the first bullet with the
following:
• Select the 400 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.

On page 2-21, under step 3, replace the first bullet with the
following:
• Select the 400 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.