114, 115, 116, and 117
Calibration Information

10

Table 3. DMM Performance Tests (cont)

Display Reading

Step Function Range Applied

114

[1]

115

[1]

116

[1]

117

[1]

18

6.000V

0V

-0.002 to 0.002

19

6.000V

5V

4.973 to 5.027

20

60.00V

50V

49.73 to 50.27

21

600.0V

600V

596.8 to 603.2

22

L

DC Volts

600.0V

-600V

-596.8 to -603.2

23

600.0 mV

6 mV, 45 Hz

5.6 to 6.4

24

m

AC Millivolts

600.0 mV

600mV, 1 kHz

587.7 to 612.3

25

600.0 mV

10 mV

9.7 to 10.3

26

DC Millivolts

600.0 mV

600mV

596.8 to 603.2

27

G

Diode

2.000 V

1.9V

N/A

1.881 to 1.919

28 1000

nF

Open

N/A

0 to 2

0 to 2

0 to 2

29

E

Capacitance

9999

μ

F 900

μ

F

N/A

881 to 919

881 to 919

881 to 919

30 LoZ

Capacitance

10.00

μ

F 9

μ

F

N/A

7.18 to 10.82

Set calibrator to standby, reconfigure leads, and program for amps output

31

A

DC Amps

10.00 A

10A

N/A

9.87 to

10.13

N/A

9.87 to

10.13

32

?

AC Amps

6.000 A

5.0A, 45 Hz

N/A

4.922 to

5.078

N/A

4.922 to

5.078

33

$

DC

μ

amps

600.0

μ

A 600

μ

ADC N/A

N/A

593.8 to

606.2

N/A

34

$

AC

μ

amps

600.0

μ

A 600

μ

AAC, 45 Hz

N/A

N/A

590.7 to

609.3

N/A

35

Open

input

N/A

N/A

0PEn

N/A

36

0.0

°

C

N/A

N/A

-1.0 to 1.0

N/A

37

y

Temperature

400

°

C N/A

N/A

395.0 to

405.0

N/A

38

0.5 V, 45 Hz

0.2 to 0.8,

AC

Annunciator

On

N/A

0.2 to 0.8,

AC

Annunciator

On

0.2 to 0.8,

AC

Annunciator

On

39

0.5 V, 0 Hz

0.2 to 0.8,

DC

Annunciator

On

N/A

0.2 to 0.8,

DC

Annunciator

On

0.2 to 0.8,

DC

Annunciator

On

40

x

500

V

[2]

,

500 Hz

489.7 to

510.3

N/A

489.7 to

510.3

489.7 to

510.3