Users Manual

18

Specifications

Infrared

Measurement Range ................................................................-40

°

C to 550

°

C (-40

°

F to 1022

°

F)

Spectral Range .........................................................................8 to 14 microns
Accuracy...................................................................................

±

1 % or

±

1

°

C (2

°

F); < 0

°

C (32

°

F),

±

1

°

C (2

°

F)

±

0.1

°

/1

°

(Assumes ambient operating temperature
of 23 to 25

°

C (73 to 77

°

F))

Repeatability.............................................................................

±

0.5 % of reading or

±

1

°

C (2

°

F)

Response Time (95 %).............................................................500 ms
Distance to Spot (D:S)..............................................................12:1
Emissivity Adjustment ..............................................................Three settings: low (0.3), medium (0.7),

high (0.95)

Contact Probe

Probe Type ...............................................................................Wrap thermocouple type K probe with

miniconnector

Measurement Range ................................................................0

°

C to 100

°

C (32

°

F to 212

°

F)

Probe Accuracy ........................................................................

±

2.2

°

C (4

°

F)

Display Resolution....................................................................

±

0.1

°

C (0.1

°

F)

Secondary Display Information ................................................Maximum, Minimum, Differential, KTC

Laser

Sighting ....................................................................................Single point laser
Power .......................................................................................Class 2 (II) operation; Output <1 mW,

wavelength 630 to 670 nm

Electrical

Power Supply ...........................................................................2 AA Batteries (alkaline or NiCD)
Power Consumption .................................................................At least 12 hours battery life

Physical

Weight ......................................................................................17.69 cm (6.965 in) H x 16.36 cm (6.441

in) L x 5.18 cm (2.039 in) W

Size...........................................................................................0.322 kg (0.7099 lb)
Total Length of Thermocouple .................................................Approximately 100 cm (40 in)

Environmental

Operating Temperature Range..................................................0

°

C to 50

°

C (32

°

F to 120

°

F)

Relative Humidity .....................................................................0 to 90 %, noncondensing up to 30

°

C

(86

°

F)

Storage Temperature................................................................- 20

°

C to 65

°

C (-4

°

F to 150

°

F)

Optional Accessories ................................................................Soft Case