Fluke 52 User Manual

Page 18

Advertising
background image

51, 52, 53, 54 Series II

Service Manual

10

Calibration
button access

zb02f.eps

Figure 4. Calibration Button Access

3.

The menu on the display changes to “

CAL A-d

”.

Note

Within

CAL

are four menu options that may be accessed by pressing

E:

CAL A-d

prompts the operator for the A/D calibration inputs in mV dc.

CAL rEF

allows the operator to calibrate the reference junction at a

known temperature.

CAL donE

saves the new calibration constants in memory.

CAL Abrt

ignores calibration changes (turning the thermometer off

while in

CAL

will also ignore calibration changes).

4.

Press

K or N to select different menu options.

5.

With

CAL A-d

selected, press

E. The thermometer prompts “80”. Set the

calibrator to +80.000 mV and allow the output to stabilize.

Note

For steps 6 and 8, keep hand and body motion to a minimum while the
thermometer takes A/D readings.

6.

Press

E. Wait while it takes A/D readings.

7.

When the thermometer prompts for “-8”, set the Calibrator to -8.0000 mV. Allow the
output to stabilize.

8.

Press

E on the thermometer and wait while it takes A/D readings. The

thermometer now shows the measured value in the bottom portion of the display.
Millivolts appears on the left side and microvolts appears on the right side.

9.

If the measurement is from –7 995.0 to –8 005.0, press

E and the meter will

return to the calibration menu and display

CAL rEF

. If the measurement is outside –

7995.0 to –8005.0, press

E once to return to the calibration menu. Using

K or N, select

CAL A-d

and try the A/D calibration procedure a second time

starting at step 5.

10.

Using

K or N, select “

CAL donE

”, then press

E to store the calibration in

memory.