EMC Compliance

This instrument meets the requirements of the EMC Directive 89/336/EEC.

Compliance was demonstrated by meeting the test limits of the following standards:

Emissions

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits
used were:

a) Radiated:

Class

B

b) Conducted:

Class

B

c)

Harmonics:EN61000-3-2 (2000) Class A

The instrument is Class A by product category.

Immunity

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test meth-
ods, limits and performance achieved were:

a)

EN61000-4-2 (1995) Electrostatic Discharge: 4 kV air, 4 kV contact

Performance A.

b)

EN61000-4-3 (1997) Electromagnetic Field: 3 V/m, 80 % AM at 1 kHz

Performance A.

c)

EN61000-4-11 (1994) Voltage Interrupt: 1 cycle, 100 %

Performance A.

d)

EN61000-4-4 (1995) Fast Transient: 1 kV peak (ac line), 0.5 kV peak
(signal lines and RS232/GPIB ports)


Performance A.

v