3

Specifications and Environmental
Conditions

3.1

Specifications

Temperature Range

35°C to 400°C (95°F to 752°F)

Display Accuracy

±0.5°C to 200°C
±1.0°C to 400°C

Stability

±0.2°C to 300°C
±0.3°C to 400°C

Resolution

0.01°

Uniformity

±0.3°C at 100°C
±0.6°C at 200°C
±0.9°C at 300°C
±1.4°C at 400°C

Heating Time

25

°C to 400°C: 22 minutes

Cooling Time

400

°C to 100°C: 65 minutes

Stabilization Time

8 minutes

Controller

Model 2200 Microprocessor based with RS-232 serial port

Readout

°C or °F switchable

Sensor

RTD, 100

Ω

Heater

325 Watt, solid state controlled

Surface Plate

6061 aluminum; top surface machine finished to 0.000032" (0.0008 mm), 3.8" (96
mm) diameter accessible

Power

115 VAC (±10%), 2.8 A or 230 VAC (±10%), 1.4 A, specify, 50/60 Hz, 325 W

Weight

7 lb. (3.2 kg) with 2200 Controller

NIST-Traceable
Calibration

Data at 50°C, 120°C, 190°C, 260°C, 330°C, and 400°C

Safety

OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1

3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. The instrument operates
safely under the following conditions:

• temperature range: 5-40°C (41-104°F)

9