2 temperature control and cutout sensor 20, Figure 3 back view 20, Temperature control and cutout sensor – Fluke 9112A User Manual

Page 22: Figure 3, Back view, 2 temperature control and cutout sensor

Advertising
background image

5.2.2

Temperature Control and Cutout Sensor

The temperature control sensor is a Type K Thermocouple as indicated. This
sensor is 3/16 inch in diameter and 12 inches long. Its location in the block is
important and can cause the gradient in the block to move back and forth. The
probe is normally inserted as shown in Figures 2 and 3.

The cutout sensor is the same as the control sensor, 12 inches long. This sensor
is inserted through a tube in the back of the block. Its location here helps pre-
vent the heater elements from overheating thus prolonging their life.

9112A Calibration Furnace
User’s Guide

20

Figure 3 Back View

Advertising