Technical specifications, Ordering information – GE ELECTROPHORESIS EPS 301 User Manual

Page 11

Advertising
background image

9

8.

Technical specifications

Regulation

Maximum voltage and current with automatic cross-over at preset limits

Output mode

Contante Voltage: 0-300 V DC,
Contante Current: 0-400 mA

Maximum Power Output

80 W

Programming range

Voltage: 5-300 V DC
Current: 10-400 mA
Time: 00.01-23.59 h or off (- - - -)

Output resolution

Voltage: 1 V
Current: 1 mA

Programming resolution

Voltage: 1 V
Current: 1 mA
Time: 1 min

Accuracy

Voltage: 4%, ± 2 V
Current: 4%, ± 4 mA
Timer: 0.1% ± 1 min

Line regulation

< 0.5%

Load regulation

< 1% at load change 10-90% of maximum load

Ripple

< 3% at 300 V

Short term stability

< 0.2% /10 h after warm up

Long term stability

< 1% /year

Output protection

Fully protected against any overload conditions

Recovery after power failure

The program continues automatically

Ambient operating temperature

4-40 °C

Ambient operating humidity

0-95%

Ambient operating pressure

68-106 kPa, maximum altitude of 2000 m

Mains requirements

100-120 V/220-240 V; 50/60 Hz

Power consumption

Max 120 W

Dimensions (WxDxH)

250 x 215 x 95 mm

Weight

3.0 kg

9.

Ordering information

Designation

Code No.

Electrophoresis Power Supply - EPS 301

18-1130-01

EPS 301 - User Manual

Advertising