PUMP: FRONT PANEL
Front panel of pump is shown below. Descriptions of but-
tons and indicators follow.

Tuning button;

indicators at left of button

Power failure indicator

Alarm mute button

Service indicator

Power (ON/OFF) button;

indicator at right of button

Panel lock button;

indicator at right of button

Maxfirm button

Auto-Adjust button;

indicators at right of button

Low pressure

 indicator

Therapy adjust button;

indicators at right of button

Front panel, first row

Auto-Adjust button: The auto-adjust button

adjusts the mattress pressure automatically based on the
patient's weight. To remind the user that the auto-adjust
process is ongoing, the auto-adjust indicator LEDs will
periodically flash from left to right until the Ready LED illu-
minates to show the process is complete. Three conditions
trigger the auto-adjust function:
1. Press the auto-adjust button for two seconds; the mat-

tress system automatically readjusts pressure setting.

2. When initial mattress inflation is complete, the mattress

system automatically starts the auto-adjust function.

3. When the mattress system detects a significant change

in patient's weight on the mattress (e.g. patient egress
or ingress), the system automatically starts the auto-
adjust function.

When auto-adjust is complete, the Ready indicator will illu-
minate. To disable the auto-adjust function, press the auto-
adjust button until auto-adjust indicators de-illuminate.
Note: During the auto-adjust operation, it is normal for the
mattress system to cycle through a series of inflation and
deflation.