0 symbol definition and button locations, Model 315 humidity temperature meter – Monarch Instrument MINI HUMIDITY TEMPERATURE METER 315 User Manual

Page 5

Advertising
background image

MODEL 315 HUMIDITY TEMPERATURE METER

3.0 Symbol Definition and Button Locations:

Feature Locations:
1) Dust

mask

2) LCD

display

3) ON/OFF

button

4) MAX MIN button
5) HOLD

button

6) RELATIVE

button

7) °C, °F control button
8) Battery

cover

2

HUMIDITY:

TEMP:

4

6

7

5

3

2

F

140

4 F

0

%RH

100

MIN

MAX

¢XF

C

20

¢XC /

C

60

REL

HOLD

8

1

HUMIDITY

OPEN

9V BAT

NEDA 1604 6F22 006P

TERY

Auto power off

indication

Temperature display

Unit indication

Data Hold indication

Minmum indication

Low battery indication

Maximum indication

Relative indication

Humidity display

MODEL 315 HUMIDITY TEMPERATURE METER

5

5.2 Humidity

Calibration

1. Turn the unit off. Press and hold the HOLD and °C/°F buttons. While

these two buttons are pressed, turn the power on. Release all buttons
and all the segments on the display will blink.

2. Within 3 seconds of releasing these buttons, press the

UREL button to

enter the calibration mode, or the instrument will revert to normal
operating mode.

3. Upon entering the calibration mode, ‘CAL1’ will appear on the second

display.

4. Place the probe in a stable 32.8% RH environment at 25°C and allow

the system to stabilize for 30 minutes. Press the MAX MIN button to
initiate calibration. A single long beep will indicate calibration is
complete, and the instrument will advance into the ‘CAL2’ mode, as
indicated on the second display.

5. Insert the probe in a stable 75.3% RH environment at 25°C and again

allow the system to stabilize for 30 minutes. Press the MAX MIN button
to initiate calibration. A single long beep will indicate calibration is
complete, and the instrument will revert to normal operating mode.

5.3 Restoring Factory Calibration

1. To restore factory default calibration values, turn the instrument off.

Press the HOLD and °C/°F buttons. While these two buttons are
pressed, turn the power on. Release all buttons and all the segments
on the display will blink.

2. Within 3 seconds of releasing these buttons, again press the HOLD

button to restore factory calibration values. A single long beep will
indicate restoration is complete.