Tray icon – Monarch Instrument 306 User Manual

Page 4

Advertising
background image

MODEL 306

DUAL CHANNEL DATALOGGING THERMOMETER

1

1.0 Introduction:

This instrument is a dual channel, recording digital thermometer for use with any Type K
thermocouple as a temperature sensor. Temperature indication complies with the NIST and
IEC584 temperature/voltage tables for Type K thermocouples. Its internal memory can store up
to 16312 records in any number of recording intervals. It is equipped with a RS232 interface for
bi-directional communication with a PC.

2.0 Specifications:

Measurement Range: -200°C - 1370°C

-328°F - 2498°F

Accuracy: (At ambient of 23 ± 5°C)

Range Accuracy

-200°C - 200°C

±(0.2% reading + 1°C)

200°C - 400°C

±(0.5% reading + 1°C)

400°C - 1370°C

±(0.2% reading + 1°C)

-328°F - -200°F

±(0.5% reading + 2°F)

-200°F - 200°F

±(0.2% reading + 2°F)

200°F - 2498°F

±(0.3% reading + 2°F)

Note:

The basic accuracy specification does not include the error of the probe. Please refer to the probe
accuracy specification for additional details.

Resolution:

-200°C - 200°C: 0.1°C;

200°C -1370°C:

1°C

-200°F - 200°F:

0.1°F;

outside this range:

1°F

Number of Sensor Inputs: Two Type K Thermocouples
Input Protection at Thermocouple Input:
60V DC or 24Vrms AC
Display:

Three 4 Digit Temperature Displays plus Indicators, see Section 3.0

Switchable

backlight

Sample Rate: 1.25 times per second
Time Function:
Real time clock, see Section 4.6
Datalogging Capacity:

16,312 records

Recording Interval:

User adjustable, see Section 4.7

Digital Output:

Bi-directional RS232; Software and Cable included

Power requirement:

9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22

Battery Life:

Approx. 100 hours with alkaline battery; Low battery indication

Operating Conditions:

x Operating Temperature and Humidity: 0°C - 50°C (32°F - 122°F); 0 - 80% RH
x Storage Temperature and Humidity: -10°C - 60°C (14°F - 140°F); 0 - 80% RH
x Altitude: Up to 2000 meters (6500 feet).

Dimensions:

184Ч64Ч30 mm (7.3Ч2.5Ч1.2 in)

Weight:

Approx. 210g (7.4 oz)

Accessories:

Two Type K Wire Thermocouple Probes, Battery, Carrying Case,

Instruction Manual, Software, RS232 Cable

Option :

AC Adapter: 9VDC ±15%; 100mA: Plug Diameter: 3.5mm×1.35mm

Temperature Coefficient:

For ambient temperatures from 0°C - 18°C
and 28°C - 50°C, for each °C ambient
below 18°C or above 28°C, add the
following tolerance into the accuracy spec.

0.01% of reading + 0.03°C

(0.01% of reading + 0.06°F)

MODEL 306

DUAL CHANNEL DATALOGGING THERMOMETER

5. Exit:

Terminates ThermoLog program.

6. Help:

On-line help for ThermoLog.

7. DataLogger:

The DataLogger window controls the loading of recorded data from thermometer.

8. Tabular:

The Tabular window displays the present data from the thermometer in a scrolling table.
Data can be stored as a file, or the table can be copied to other software, such as EXCEL,
for further analysis.

9. Tray Icon:

When ThermoLog is running, an icon appears in the Windows Tray area (see figure below),
to display a pop-up menu.

Tray Icon

Advertising