SECTION 1—GENERAL GUIDELINES

Supine Bath Tub

10

Part No 1096926

When operating water pressure exceeds 55 p.s.i., pressure regulators (similar to the 
¾‐inch Watts U5) MUST be installed.

Maintenance and Inspection

If any part of the system is NOT functioning properly, DO NOT use the system. Have the 
system serviced to correct the problem.

The system MUST be maintained on a regular basis to ensure it is functioning properly.

Electrical Shock Hazard

All disassembly and maintenance of the tub MUST be done by a qualified technician, 
certified electrician or plumber. Disconnect electrical supply or turn Off circuit breaker 
before performing any maintenance to the tub.

Disinfecting

Disinfect the tub after initial set up and ALWAYS disinfect the tub AFTER each use to 
avoid resident infection and contamination of the tub.

DO NOT disinfect the tub when it is occupied otherwise injury may occur.

ALWAYS wear gloves, protective apron, and face shield protection when disinfecting or 
cleaning your tub.

ALWAYS handle disinfectants, cleaning compounds, and powdered sanitizers with 
extreme care. Physical or mechanical damage caused by improper handling or usage of 
such products will not be the responsibility of ICCG.

DO NOT attempt to turn on whirlpool pump during the disinfectant cycle. Doing so 
could result in skin or eye contact with the disinfectant, which is corrosive. If contact is 
made, rinse IMMEDIATELY with water. If irritation occurs, consult a physician. 

ALWAYS read and follow disinfectant label directions carefully.

NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice 
stone. The use of these types of cleaners will make the gel‐coat finish of your tub scratched 
and dull. NEVER use cleaners containing iodine, bromine, Betadine

®

 or methylene blue. 

The use of these types of cleaners will cause the tub to stain. Never use cleaners containing 
bleach. Use of these types of cleaners will dry out the rubber seals and gaskets and the tub 
will not function properly.

Operation

DANGER

Keep hair and body away from suction guard when pump is running.

Advertising