Troubleshooting – Invacare IRC1175 User Manual

Page 21

Advertising
background image

21

TROUBLESHOOTING

PROBLEM

POSSIBLE SOLUTIONS

Unit does not run from battery.

1. Assure battery is installed

properly.

2. Remove Protective film (tape).

3. Check all leads are removed

from DC inlet.

4. Charge battery for 16 hours.

Unit does not charge and

1. Confirm power pack is plugged

GREEN Charging indicator is

into live wall outlet.

NOT illuminated.

2. Have Freeway Freedom serviced

by your Dealer.

Unit does not run from DC inlet.

1. Check fuse in DC cigarette

lighter lead.

2. Check for correct polarity (Refer

to SPECIFICATIONS Section

in this manual).

Unit runs slowly or battery

1. Check if filter is clean. Replace.

usage time decreased.

2. Check coiled duratube for

blockage. If blocked, replace

with new one. DO NOT

immerse in water.

3. Follow Discharge/Charge

directions in"Prolonging the Life

of your Battery". (Refer to

MAINTENANCE Section in

this Manual.) Procedure may

need to be repeated.

Unit runs and Sidestream

1. Thoroughly clean the nebulizer

nebulizer makes poor or no mist.

(Refer to AFTER THE

TREATMENT section in

this manual.)

2. Replace nebulizer with a new

one

.

NOTE: If any of the above problems occur and the solutions do not work,

please contact your Invacare Dealer for service.