SECTION 4—TYPICAL PRODUCT PARAMETERS

Platinum

Series

14

Part No 1118353

Flow Range:

Platinum 5 and XL - 0.5 to 5 L/min For
flowrates less than 1 L/min, we recommend
the use of the Invacare Pediatric Flowmeter
IRCPF16 Accessory.

Platinum 10 - 2 L/min to 10 L/min

Low Flow Alarm:

Platinum 5 - 0 L/min to 0.5 L/min,
Platinum 10 - 0 L/min to 1 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected)

Average Power
Consumption:

Platinum 5 and XL: 400W
Platinum 10: 585W

Pressure Relief Mechanism
Operational at:

35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)

Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:

0.7 L/min

Filters:

Cabinet (2), Outlet HEPA and Compressor
Inlet.

Safety System:

Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO

2

Oxygen System (SensO

2

Model)

Low Flow Alarm.

Width:

18 3/8 in ± 3/8 in (46.7 cm ± 1 cm)

Height:

26 3/8 in ± 3/8 in (67.0 cm ± 1 cm)

Depth:

14 3/8 in ± 3/8 in (36.5 cm ± 1 cm)

Weight:

Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ±
1 kg)

Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg)

Shipping Weight:

Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ±
1 kg)

Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg)

Advertising