SECTION 7—MAINTENANCE

Part No. 1078984

23

Stand Up Patient Lift

Replacing the Electric Actuator

1.

Remove the bottom nut, washer and shoulder bolt that secure the electric actuator to the mast mounting bracket.

2.

Rest the lift arm on your shoulder and remove the top nut, bolt, bracket and bushing from the lift arm mounting bracket.

3.

Remove the electric actuator.

4.

Reverse STEPS 1‐3 to install the new electric actuator.

ƽ WARNING

DO NOT overtighten mounting hardware. This will damage mounting brackets.

FIGURE 7.3 Replacing the Electric Actuator

Bolt

Washer

Bushing

Washer

Bolt

Bolt

Washer

Bushing

Washer

Bolt

Bracket