Troubleshooting display codes – Invacare RLS6 User Manual

Page 7

Advertising
background image

Part No 1114832

7

Reliant Scale RLS6

7. Press the LOCK/UNLOCK key. The display reads UNLD.

8. Press the ZERO key. The display reads LOAD.

9. Hang 200 pounds of calibrated weight from the scale and press the ZERO key.

10. Press the ZERO key. The display reads DONE.

11. Remove the weight. Press the LOCK/UNLOCK key.

NOTE: The scale is now in test mode.

12. Press the ZERO key. The display reads 0.00 lb.

NOTE: The scale will now be in normal operation.

FIGURE 5 Calibrate the Reliant Scale

Troubleshooting

Display Codes

Calibration Required ‐ Indicates improper stored calibration data, calibration is necessary.

Over Capacity ‐ Indicates a weight exceeding the capacity has been loaded on the scale.

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

Unit does NOT work properly.

Battery failure.

Check battery. Replace if necessary.

Battery has been replaced and unit
still does NOT work properly.

Contact Invacare for Service at
1-800-333-6900.

CAL Button

BACK VIEW OF THE SCALE DISPLAY

FRONT VIEW OF SCALE DISPLAY

Display Window