I.R.I.S. FLAME MONITORING SYSTEM P522 User Manual

Page 22

Advertising
background image

IRIS MODEL P522 APPLICATION MANUAL

Page 22

you should definitely turn off the Si channel, which
cannot contribute to the flame signal from the burner
(when it is gas) but can only read background glow –
which will certainly cause a discrimination problem.

An important aspect of using the S512 instead of the
S509 has to do with linearity. The non-linearity of
the Si channel due to the AGC circuit has already
been explained. This AGC is used upstream of the
25-turn potentiometer, so you have no control over
this action except for the viewing head sighting, or
for using orificing to change the amount of signal
reaching the photodetector. On the other hand, the
PbS channel upstream from its potentiometer is lin-
ear for low to medium brightness flames (i.e., no
distortion of the flame signal occurs with regard to
amplifier saturation (signal clipping) or nonlinear
amplification (AGC) as with the Si channel).

However, saturation will occur with the PbS channel
on high brightness flames. You can detect this by
observing the digital display reading for boiler load
conditions as they vary from low to high. If there is
not much of a signal increase between low load and
high load, then the PbS detector is saturating.

For example: if the burner flame signal
on oil reads 1950 under low load condi-
tions, and reads 2000 on high load con-
ditions, you can be sure that the PbS de-
tector is saturating.

To resolve this condition, reduce the signal bright-
ness by orificing (see next section ORIFICING).
The orifice disc (installed in the flange) is located
immediately in front of the lens and will reduce the
amount of signal radiation the same way a camera
aperture reduces the amount of light.

As mentioned in the previous section on the S509,
saturation can occur downstream from the gain ad-
justment potentiometers because of the signal ampli-
fication process. This type of saturation can be de-
tected on the digital display by observing the signal
count. The S509 will be completely saturated at a
count of 2800 – 3000. Saturation begins (waveform
just starts to clip) at a count of 2100. This threshold is
somewhat dependent upon the flicker frequencies and
their distribution, but for the lower frequencies (up to

200 Hz) the count of 2100 can be taken as the thresh-
old. This amplifier type saturation usually will not
occur with the S512 viewing head when reading ac-
tual flames. It can occur when reading artificial sig-
nal sources, such as incandescent and fluorescent
lamps, because of the large AC/DC component. It
should be noted that amplifier type saturation is not
harmful per se, but saturation of the photodetector
caused by high brightness flames can be.

You cannot control the nonlinearity caused by the
AGC action of the Si channel on the S509, but with
the Si channel turned off you can control the linear-
ity of the PbS channel by observing the count on the
digital display, and reducing the gain when neces-
sary by orificing. Saturation of the PbS detector can
only be corrected by orificing; turning down the gain
adjustment potentiometer will not help, because it is
the PbS cell upstream that is the cause.

Another nonlinearity of the PbS detector that should
be noted has to do with the signal sensitivity and
temperature. The PbS-type detector becomes less
sensitive with increasing temperature.

For example: the sensitivity or signal will
drop to one half with an increase in tem-
perature from 25°C to 50°C. A further
increase in sensitivity will occur above
50°C. This is the reason for the 60°C
operating temperature limitation on the
viewing head. This characteristic is in-
herent in all PbS detectors, but should not
cause a problem with respect to flame
monitoring and discrimination. The sig-
nal level will be stable once a tempera-
ture equilibrium has been reached.

This sensitivity-versus-temperature characteristic is
another reason why the PbS cell should not be ex-
posed to high brightness flames. The radiation of the
flame is focused on the PbS element, causing its tem-
perature to rise, reducing the displayed count. This
reduction in output with increased flame intensity is
another form of saturation. Also, the exposure to el-
evated temperatures over a period of time causes an
aging process resulting in a change of the dark resis-
tance of the PbS element, and a loss of sensitivity.

Advertising