Wall Mount Assembly

EASYTRACK WALL MOUNT KIT
CONTENTS & PART NUMBERS (Figure 8)

Description

Part #

Qty

1

Wall mount plate

900-1085

1

2

Pivoting rail
attachment

900-1087

1

Includes:
3

Round blue cap

900-1072

1

4

Pivot wall mount

N/A

1

5

Fixation wall mount

N/A

1

6

Back plate

N/A

1

7

M 10 X025 screw

900-0409

1

ASSEMBLY OF WALL MOUNT ASSEMBLY

Tools required:

Step ladder

Tape measure

6 concrete 3/8” anchors or 6 3/8” lag bolts

Drill

Wrench with a 9/16” socket

Someone to assist you

Stud finder

1. Refer to Figures 9A and 9B for the distance required between the ceiling and where

the top of the wall mount must be installed. Please note that when the wall bracket
will have the pivoting rail of a 3 post system attached, the mounting point must be
lower than with a standard rail.

2. Insert the pivoting rail attachment into the wall plate. Refer to Figure 10 to insure

that the bracket is placed properly.

3. Install the wall plate on the wall. The wall plate must be installed on a concrete wall

with six (6) 3/8 in. concrete anchors or on three (3) 2x3 (minimum size) wooded studs
with two (2) 3/8" diameter (M10), 4 in. long lag bolts on each stud (Figure 11).
Wood studs– Use a stud finder to find the center of each stud. It is necessary to pre-
drill holes into each stud before screwing the lag bolt to prevent the wood from
cracking.

NOTE– Make sure the rail is level before securing wall mount to wall.

Follow drill size recommendation for lag bolt:

LAG BOLT

M10

3/8 "

PRE-DRILL SIZE

6.5 mm

1/4 "

Pre-drilled holes must be in the center of the stud. If holes are not properly
centered, serious damage or injury may occur.

DO NOT install into steel stud, reinforcement may be required.

CAUTION– Avoid drilling close to any electrical wiring or plumbing. Please, make sure you know what

is behind the wall you are drilling in before doing any modifications to it. Failure to do so
may cause serious injury.

4. See “Before you start” for suggested position. Slide the pivoting rail attachment into its

final position. For the 3 post assembly, the bracket must be placed in a way that the rail
won’t, when turning, generate an angle beyond the maximum allowed (Figure 12).

5. Tighten the four bolts on pivoting rail attachment. The rail can now be attached to

wall mount.

A s s e m b l y I n s t r u c t i o n s

5

8

9a

9b

10

11

12

standard rail

pivoting rail

1

2

3

4

5

6

7

ceiling

bottom

30°

30°

Maximum angle

1 1/8"
(28mm)

5 1/4"
(133mm)

Advertising