Control, alarm and monitoring, Twin guard series, 86°c ultra-low freezers – Sanyo MDF-U500VXC User Manual

Page 9

Advertising
background image

9

Control, Alarm and Monitoring

Twin Guard Series

-86°C Ultra-Low Freezers

Alarm & Safety Features

Events

Visual

Audible

Signal to Alarm Contact

Status Alert

Abnormal ambient (too high or

too low), or abnormal freezer

loading (too much warm

product at one time).

Notification of error on

graphical display

None

None

High Temperature (+40°C)*

Interior chamber warms

beyond high temp setpoint

Intermittent tone. Time delay

of 15 minutes after reaching

alarm setpoint avoids false

alarms. Intermittent tone.

Yes

Low Temperature (-40°C)*

Interior chamber warms

beyond low temp setpoint

Yes

Power Failure

Loss of power

Intermittent Tone

N/A

Control Temperature

Sensor Failure

Sensor probe disconnect

short circuit or failure

Solid Tone

No

Auto Return

Touch key is not pressed

for 90 minutes

Reverts to chamber

temperature display

None

No

Door Alarm

Door open

Notification of error on

graphical display

Solid tone sounds after

2 minutes

Yes

*New extended range is important for GxP applications to reduce potential incident event reporting.
The integrated microprocessor control aggregates information from the entire system and compares to normal or setpoint values. Out-of-compliance events are

announced by visual and audible warnings, and remote alarm when connected to an external system.

All performance attributes are displayed on a large, visually intuitive alpha-

numeric display.

Internal temperature log files are expressed in graphical form over time.

Out-of-compliance events are announced by visual and audible warnings.

Error codes quickly identify problem sources and permit fast diagnosis

and remediation.

Advertising