Operating instructions, Alarm temperature setting – Sanyo MDF-593C User Manual

Page 18

Advertising
background image

17

OPERATING INSTRUCTIONS

Alarm temperature setting

This unit is provided with the high temperature alarm. The setting of high temperature alarm is 10

o

C or

15

o

C higher than the setting of chamber temperature. The procedure in table 2 shows the sequence to

set the high temperature alarm at 15

o

C higher than the setting of chamber temperature. Set the figure to

000 at step 5 when setting the high temperature alarm at 10

o

C higher than the setting of chamber

temperature.

Note:

The alarm temperature is set at the factory 10

o

C higher than the setting of chamber temperature.

Table 2. High temperature alarm setting (15

o

C higher than the setting of chamber temperature)

Description of operation

Key operated

Indication after operation

1

-----

The current chamber temperature is

displayed.

2

Press the numerical value shift key

for about 5 seconds.

The first digit blinks.

3

Set the first digit to 1 with the

numerical value shift key.

The first digit blinks.

4 Press ENT key.

ENT

The first digit blinks.

5

Set the first digit to 1 with the

numerical value shift key.

When pressed, the figure of settable

digit changes.

6 Press ENT key.

ENT

Alarm temperature is memorized and

the current chamber temperature

is displayed.

Note: The setting is not memorized if the mode change key (PV/SV) is pressed during alarm

temperature setting. The chamber temperature is displayed for 3 seconds when the numerical value

shift is pressed 5 times continuously.