SENSORS

Electrochemical

Electrochemical sensors are fuel cell-like devices consist-
ing of an anode, cathode, and electrolyte. The compo-
nents of the cell are se-
lected so a subject gas
allowed to diffuse into the
cell will cause a chemical
reaction and generate a
current. The cells are dif-
fusion limited, so the rate
at which the gas enters
the cell is solely depen-
dent on the gas concentration. The current generated is
proportional to the rate of consumption of gas in the cell.

Unlike other electrochemical sensors where gas diffuses
through membranes into the cell, Sierra Monitor sensors
allow gas to diffuse into the sensor through a simple cap-
illary. The result is an extremely stable sensor with very
low temperature and pressure coefficients.

Common Specifications for All 5100 Series Sensors

Communication to Controller: SentryBus PSG
Zero Drift:

Less than 5% per year

Calibration Frequency:

90 Days Recommended

Transmission:

Digital – RFI, EMI immune

Warranty:

2 years

Housing:

Explosion proof (NEMA 7)

(Div. 1, Class 1, Groups C,D)

Optional upgrade to Group B,
NEMA 4X

Dimension:

7.6 X 4.6 X 4.2 inches
(19.3 X 11.7 X 10.7 cm)
Note that 5100-26 is larger

Weight:

2.7 lb. (1.3 Kg)

GAS SENSOR SPECIFICATIONS

Gas

Model

Sensor

Range

Time

Sensor

Operating

Relative

Number

Type

Response

Life

Temperature

Humidity

Carbon Monoxide

5100-16

Electrochem 0-2,000 PPM

90% <35 sec. 2 yr.

4

o

to 122

o

F (-20

o

to 50

o

C)

15-99% RH

(Hydrogen Tolerant to
2,000 PPM H2)
Nitric Oxide

5100-19

Electrochem 0-20 PPM

90% <15 sec. 3 yr.

4

o

to 122

o

F (-20

o

to 50

o

C)

15-99% RH

Hydrogen Cyanide

5100-22

Electrochem 0-20 PPM

90% <70 sec. 2 yr.

4

o

to 122

o

F (-20

o

to 50

o

C)

15-99% RH

Ethylene Oxide

5100-27

Electrochem 0-20 PPM

90% <90 sec. 2 yr.

4

o

to 122

o

F (-20

o

to 50

o

C)

15-99% RH

Sierra Monitor Corp. 1991 Tarob Ct., Milpitas, CA 95035, USA, 408-262-6611, 800-727-4377, FAX: 408-262-9042