ASSEMBLY INSTRUCTIONS (FIGURE 1)

1. Remove the mattress from the bed.
2. Place the bed in the flat position.
3. Measure the width of the bed frame. Attach the two crossbrace assemblies

together.

4. Place the assembled crossbrace on the bed frame. Attach clamps as shown and

tighten wing nuts until secure. Prevent the lock nut from contacting the bed
frame.

5. Once the crossbrace is adjusted properly, tighten the center bolts and nuts.
6. Repeat steps 3-5 for other crossbrace. Ensure the other crossbrace is between

62" to 64" apart as shown.

7. The slide tube can be rotated with the locking knob on the right or left by

rotating the slide. The bed rail can be adjusted higher or lower based on your
rail height preference.

8. Insert the rail into the slide tubes. Pull the plunger knob to adjust the rail

height.

RISK OF SERIOUS INJURY OR DEATH. When using side rails, use a mattress thick
enough and wide enough so that the gap between the top of the mattress and
the bottom of the side rails and the gap between the side of the mattress and
the side rails is small enough to prevent a patient from getting his or her head
or neck between the mattress and the side rail. Assure that articulating the
bed, head section or foot section does not create any hazardous gaps. Failure
to do so could result in injury or possible suffocation.

RISK OF SERIOUS INJURY OR DEATH. When using side rails, only use side rails
that can be positioned so that the gap between the headboard or foot board
and the side rails is small enough to prevent a patient from getting his or
her head or neck between the mattress and the side rail or large enough that
the trunk and hips can easily pass through it. Assure that articulating the
bed, head section or foot section does not create any hazardous gaps. Failure
to do so could result in injury or possible suffocation.

RISK OF SERIOUS INJURY OR DEATH. When using side rails, only use side rails
in which any gaps between the rail body members are small enough to prevent
a patient from getting his or her head or neck inside the rail body. Failure to
do so could result in injury or possible suffocation.

RISK OF SERIOUS INJURY. When using multiple side rails on one or both
sides of the bed, only use side rails that can be positioned so that the gap
between the side rails is large enough that the trunk and hips can easily
pass through it. Assure that articulating the bed head section or foot section
does not create any hazardous gaps.