A Check to see if the batteries have

been correctly inserted into the com-
partment on the underside of the
scales. Take note that the batteries
must be inserted in accord with the
polarities marked in the compartment.

B Check to see if the batteries in the

scales are discharged. Before the
batteries are completely discharged
a warning symbol is shown in the
display

ᕡ. This is also explained in

chapter 10. „Maintaining the scales“.
Exchange both button batteries.

11.2. Malfunctions in the display
The scales contain sensitive electronic
components. It is thus possible that they
may be disturbed by radio transmitter
appliances in the immediate vicinity.
These could be for example mobile tele-
phones, walkie-talkies, CB radios, radio
remote controls/other remote controls
and microwave appliances. Should mal-
functions appear in the display

ᕡ, remove

all such appliances from the vicinity of
the scales. Frequently it is also sufficient
to simply switch off the particular app-
liance as soon as the scales are to be
taken into use.

NOTE:

Avoid the loading and discharging of
static electricity. This can cause erroneous
measurement displays and/or sudden
inactivation of the scales.

11.3. Error messages shown in the

display

11.3.1. „0-Ld“
If this error message appears, the weight
on the scales exceeds 160 Kg. The scales
are not designed for weights above
160 Kg.

ƽ

Warning!

Do not burden the scales with a weight
of more than 160 kg! Through this, the
scales could be irreparably damaged!

11.3.2. „Err“
If this error message is shown the scales
cannot measure the bodyfat, muscle
mass and body-water shares.

A The measurement does not function if

shoes or socks are being worn. Remove
your shoes and socks before standing
on the scales.

B Ensure that you are standing with

both feet on the sensors on the upper
side of the scales.

Should the measurements still not be
possible:

C Step down from the scales and wait

until it automatically switches itself
off. Switch the scales back on and
start the measurements again.

Should the measurements still not be
possible: