Laser:display:set, Laser:enable:cond – Lightwave Communications LDX-3232 User Manual

Page 70

Advertising
background image

C O M M A N D R E F E R E N C E
LDX-3232 Device-Dependent Command Reference

54

LDX-3232

C H A P T E R

4

The LASer:DISplay:SET? query returns the status of the setpoint display mode.

Syntax Diagram

- where the response is 0 or 1.

Notes

The setpoint display will not time out when REMOTE operation is used. (It will be
continuously displayed.)

Examples

"LAS:DISPLAY:SET?" -response: 0, means the setpoint is not shown on the display.

"Las:Dis:Set?" -response: 1, means the setpoint value is shown on the display.

The LASer:ENABle:COND command sets the condition status enable register of the
LASER operations for summary (in bit 3 of the status byte) and generation of service
requests.

Syntax Diagram

Parameters

An <nrf value> whose sum represents the enabled bits:

Notes

The enabled or disabled LASER conditions can be read by using the
LASer:ENABle:COND? query.

The LASER condition status can be monitored by the LASer:COND? query. If any of the
enabled LASER conditions are true, bit 3 of the status byte register will be set.

The enable registers normally retain their values at power-up (as they were at power-
down) unless the power-on status clear flag is set true (for more information see the
*PSC definition, in the GPIB-IEEE488 specification).

LASer:DISplay:SET?

F

RONT

P

ANEL

R

EMOTE

LASer:ENABle:COND

F

RONT

P

ANEL

R

EMOTE

1

LASER Current Limit

256

Output is shorted

2

LASER Voltage Limit

512

Output is outside tolerance limit

4

N/A

1024

Output on/off state

8

Power limit

2048

Ready for calibration data state

16

Interlock disabled

4096

N/A

32

N/A

8192

N/A

64

Compliance Voltage Limit

16384

N/A

128

Open circuit

32768

N/A

:

LAS

:

DIS

SET

:

?

:

LAS

<white

space>

<nrf

value>

:

ENAB

:

COND

Advertising