How ua-767 plus works, Before you start, Selecting the correct cuff size – LifeSource UA-767 Plus User Manual

Page 5

Advertising
background image

USE SMALL CUFF

USE LARGE CUFF

PROPER FIT RANGE

USE SMALL CUFF

USE LARGE CUFF

PROPER FIT RANGE

INDEX

INDEX

Using the correct cuff size is important for an accurate reading. A cuff that is

too large will produce a reading that is lower than the correct blood pressure;

a cuff that is too small will produce a reading that is higher than the correct

blood pressure. With your arm hanging at the side of your body, measure the

circumference of your upper arm at the midpoint between the shoulder and

elbow.

Cuff Size Indicator On The Cuff

Our cuff has an indicator which tells you whether you are using the correct

cuff size or not. Place the cuff on your arm (see Page E-6 to learn how to put

the cuff on correctly) and if the Index Mark points within the Proper Fit

Range, you are using the correct cuff size. If the Index points outside of the

Proper Fit Range, contact A&D Medical at 1-888-726-9966 (in Canada, call

Auto Control Medical at 1-800-461-0991) for more information on where to

obtain a cuff replacement.

You must install 4 type AA (1.5 volt) batteries (alkaline batteries recommended),

or use the AC Adapter (see page E-11 for using the Monitor with AC Adapter)

and attach the cuff to the monitor before using it. To install batteries (or replace

them if the “Low Battery” symbol appears on display), proceed as follows:

1.

Remove battery compartment cover by gently pushing down on arrow

and sliding cover forward.

2.

Put in bottom row of batteries first. Place the

batteries in compartment with positive (+) and

negative (–) terminals matching those indicat-

ed in the compartment. Be sure batteries make

contact with compartment terminals.

3.

Replace cover by sliding it into the compart-

ment and gently pressing into place.

NOTE: Rechargeable batteries will not work
with this monitor.

LifeSource® blood pressure monitors are easy to use, accurate and digitally

display full measurement readouts. Our technology is based on the “oscil-

lometric method” – a noninvasive blood pressure determination. The term

“oscillation” refers to any measure of vibrations caused by the arterial pulse.

Our monitor examines the pulsatile pressure generated by the arterial wall as

it expands and contracts against the cuff with each heartbeat.
The cuff is inflated until the artery is fully blocked. The inflation speed is

maximized and the pressure level is optimized by the device. The monitor

takes measurements while the cuff is both inflating and deflating. This results

in a faster measurement providing greater comfort to the user.

+

-

+

-

+

-

+

-

OPEN

HOW UA-767 PLUS WORKS

?

BEFORE YOU START

LifeSource Health Line (Toll-Free): 1-888-726-9966

E-4

?!

ARM SIZE

RECOMMENDED
CUFF SIZE

REPLACEMENT
CUFF MODEL#

6.3” - 9.4” (16-24cm)

9.4” - 14.2” (24-36cm)

14.2” - 17.7” (36-45cm)

Small Cuff

Medium Cuff

Large Cuff

UA-279

UA-280

UA-281

In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991

E-5

!

SELECTING THE CORRECT CUFF SIZE

1932-1AIc UA787EXC•06 illust•02p name.FH7

INDE

X

AC Adapter Jack

Memory Button

START Button

Air Connector Plug

EasyCuff™ Feature

Air Hose

Battery Compartment

Battery Cover

Systolic Pressure

Diastolic Pressure

Air Socket

TM

Pulse Rate

Advertising