E Series User Manual

Connecting Signals to the NI DAQPad-6015 BNC

Analog Input

You can use each analog input BNC connector for one differential signal or
two single-ended signals.

Differential Signals

To connect differential signals, determine the type of signal source you are
using: a floating signal source or a ground-referenced signal source. Refer
to the

Differential Connection Considerations

and

Connecting Analog

Input Signals

sections of Chapter 2,

Analog Input

, for more information on

connecting analog input signals and differential connection considerations.

To measure a floating signal source, move the switch to the FS position. To
measure a ground-referenced signal source, move the switch to the GS
position. Figure A-7 shows the source type switch locations on the front
panel of the BNC DAQPads.

Figure A-7. BNC DAQPads Front Panel

Figure A-8 shows the analog input circuitry on BNC DAQPads. When the
switch is in the FS position, AI x – is grounded through a 0.1

μF capacitor

in parallel with a 5 k

Ω resistor.

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

FS

GS

1

3

5

7

9 11 13 15 17 19 21 23 25 27 29

2

4

6

8 10 12 14 16 18 20 22 24 26 28 30

USER 2

USER 1

PWR

COM

CTR 0

OUT

EXT
REF

DAC 0

OUT

ACH 6

ACH 4

ACH 2

ACH 0

ACH 1

ACH 3

ACH 5

ACH 7

DAC 1

OUT

PFI 0/

TRIG 1

FS

GS

FLOATING SOURCE

GND REF SOURCE

FS

GS

Source Type

Switch

Source Type

Switch

FS = Floating Source
GS = Grounded Source