Appendix A

Specifications

© National Instruments Corporation

A-13

NI 6115/6120 User Manual

Voltage Output

Ranges .................................................... ±10 V

Output coupling...................................... DC

Output impedance .................................. 50

Ω ±5%

Current drive .......................................... ±5 mA

Output stability....................................... Any passive load

Protection .............................................. Short-circuit to ground

Power-on output voltage
(before software loads calibration values)

NI 6115 ........................................... ±400 mV

NI 6120 ........................................... ±80 mV

Initial power-up glitch

Magnitude ....................................... ±2 V

Duration .......................................... 200 ms

Table A-5. NI 6115 Analog Output DC Accuracy Information

Nominal

Range at

Full Scale

(V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

(mV)

Temp

Drift

(%/°C)

Absolute

Acc. at Full

Scale (mV)

Theoretical Resolution

(mV)

24 Hrs

90 Days

1 Year

±10

0.0437

0.0445

0.0454

8.9

0.0006

13.5

4.88

Table A-6. NI 6120 Analog Output DC Accuracy Information

Nominal

Range at

Full Scale

(V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

(mV)

Temp

Drift

(%/°C)

Absolute

Acc. at Full

Scale (mV)

Theoretical Resolution

(µV)

24 Hrs

90 Days

1 Year

±10

0.0460

0.0468

0.0477

1.9

0.0006

6.7

305.2