Pressure variation – ResMed VPAP III & III ST User Manual

Page 42

Advertising
background image

36

IEC 60601-1 Classifications
Class II (double insulation)
Type CF

Table 7: Displayed values

* Results may be inaccurate in the presence of leaks.

** The displayed values are estimates. They are provided for trending purposes only.

Pressure Variation

Value

Range

Accuracy

Display Resolution

Pressure sensor at air outlet

Pressure

-5 to 30 cm H

2

O

±0.5 cm H

2

O

0.1 cm H

2

O

Flow sensor in flow generator*

Leak

0–120 L/min

**

1 L/min

Tidal volume

100–3000 mL

**

1 mL

Respiratory rate

6–60 BPM

±0.5 BPM

0.1 BPM

Minute ventilation

0.6–60 L/min

**

0.25 L/min

0.00

5.00

10.00

15.00

20.00

25.00

0

1

2

3

Breaths

Pressure (cm H

2

O)

10 BPM

15 BPM

20 BPM

Max Pressure

2/3 Max Pressure

1/3 Max Pressure